Machine data
measurement range
Measurement uncertainty
MPEE [μm]
MPESD = 5.0µm + (L/40mm) in accordance with VDI/VDE 2630 sheet 1.3 (without maximum value limit)
temperature range
20°C +/-2 K
In accordance with DIN EN ISO 10360-2
fitment
controller
CT master computer, CT reconstruction calculator
Software option 1
WM | Quartis 4 CT; WM | eXact Control; WM | eXact Volume; WM | eXact Analysis Standard Package
description
Description/ Description
- Ergonomic desktop CT with integrated control cabinet
- High power of 130 kV with maintenance-free radiation source/High power of 130 kV with maintenance-free radiation source
- Easy operability and high performance of application software/Easy operability and high performance oft he application software
- Quick set-up of workpieces using integrated video cameras and special software features
- High-resolution detector for maximum detail detectability
- Radiation protection in accordance with full protection device in accordance with StrSchV /Full radiation protection chamber based on German RÖV regulation
- Multi-function signal light /Multifunctional system status signal lamp
highlights
- Comprehensive software package/Comprehensive software package
- Measurement chamber video transmission
- Vibration damping of the measurement chamber (elastomer)
machine profile
Operating temperature: 18°C — 30°C 1) Relative Humidity/Relative Humidity: 40% - 60%
1) A smaller temperature range is required for metrological applications. For metrological applications a smaller temperature range is required